- Stainless steel infusion rod with facility to mount Syringe/infusion pump
- Lower shelf below baby tray to facilitate under surface phototherapy, Facility of 3 drawers for storage
- Height : 74.5″ (1890 mm), Width : 28.0″ (710 mm), Length : 39.0″ (990 mm)
- Bed to Floor Height : 37.5″ (950 mm)
- Distance between baby bed and heater: 33″ (840 mm)
- Mattress size: 30″ (762 mm) x 20″ (508 mm)
- Height of side panel above baby bed: 5.75″ (145 mm)
Performance/Electrical Specifications :
- Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
- LED Display of Skin Temp.,Air Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
- LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
- Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
- Modes of Operation: Skin Servo, Air, Manual and Pre-Warm
- Keypad lock facility
- Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
- Air mode: Temperature range: Room temp. to 39°C , Resolution : 0.1°C
- Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
- Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode) Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
- Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
- Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
- Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm Condition | Alarm Messages on LCD |
---|---|
Skin Sensor is unplugged /Faulty | Skin Sensor Fail |
Air Sensor is unplugged /Faulty | Air Sensor Fail |
Measured skin temperature higher than set temperature by 1.0°C | Skin Temp. High |
Measured skin temperature lower than set temperature by 1.0°C | Skin Temp. Low |
Measured skin temperature higher than 38.0°C (With cut-off) | Skin Over Range |
Measured skin temperature lower than 34.0°C | Skin Under Range |
Measured air temperature higher than set temperature by 1.5°C | Air Temp. High |
Measured air temperature lower than set temperature by 3.0°C | Air Temp. Low |
Measured air temperature higher than 39.0°C (With cut-off) | Air Over Range |
Heater power is 100% for more than 15mins | Chk Sens Position |
Heater is unplugged or faulty | Heater Open |
Heater Power drive fail/uncontrollable | Chk. Pwr. Drive |
Power Fail Alarm: Audio with LED indicator |