Infant Radiant Warmer

Infant Radiant Warmer

Sold: 0

Standard Specifications for NIC201a

  • Continuous head up/down facility
  • X-Ray cassette holder
  • Collapsible/lockable Side Supports made up of Unbreakable polycarbonate material
  • Slots for sensor and ventilator tubing
  • Side tray for placement of monitor
  • Stainless steel infusion rod with facility to mount Syringe/infusion pump
  • Lower shelf below baby tray to facilitate under surface phototherapy, Facility of 3 drawers for storage
  • Height : 74.5″ (1890 mm), Width : 28.0″ (710 mm), Length : 39.0″ (990 mm)
  • Bed to Floor Height : 37.5″ (950 mm)
  • Distance between baby bed and heater: 33″ (840 mm)
  • Mattress size: 30″ (762 mm) x 20″ (508 mm)
  • Height of side panel above baby bed: 5.75″ (145 mm)

Performance/Electrical Specifications :

  • Microprocessor based temperature controller with Pre-warm, Manual and Skin Servo modes
  • LED Display of Skin Temp.,Air Temp., Set Temp. (Resolution 0.1°C) & Heater Power %.
  • LCD to display modes of operation, Alarm messages, Skin Temperature in °F and Timer.
  • Thermistor based sensor with Accuracy of +/- 0.2°C (Temp. Calibration not require)
  • Modes of Operation: Skin Servo, Air, Manual and Pre-Warm
  • Keypad lock facility
  • Skin mode: Temperature range: 34 to 38°C , Resolution : 0.1°C
  • Air mode: Temperature range: Room temp. to 39°C , Resolution : 0.1°C
  • Manual mode: Adjustable Heater Power 0 to 100% (with an increment of 5%), Time duration:15mins
  • Pre-warm mode: To pre-warm the cradle before baby is placed on the bed. (No alarm will be activated during Prewarm mode) Adjustable Heater Power 0 to 25% (with an increment of 5%), Time duration:30mins
  • Overhead heater box: Swivelling, Consists of Ceramic heater (650 Watts) placed in parabolic reflector with a safety grill & Halogen observation light (12V/50W) for observing the baby
  • Operating Voltage: 220-240 VAC, 50Hz, Power: 800W (Max.)
  • Alarm: Audio alarm with 10 minute mute facility & Visual alarms with message on LCD as follows
Alarm ConditionAlarm Messages on LCD
Skin Sensor is unplugged /FaultySkin Sensor Fail
Air Sensor is unplugged /FaultyAir Sensor Fail
Measured skin temperature higher than set temperature by 1.0°CSkin Temp. High
Measured skin temperature lower than set temperature by 1.0°CSkin Temp. Low
Measured skin temperature higher than 38.0°C (With cut-off)Skin Over Range
Measured skin temperature lower than 34.0°CSkin Under Range
Measured air temperature higher than set temperature by 1.5°CAir Temp. High
Measured air temperature lower than set temperature by 3.0°CAir Temp. Low
Measured air temperature higher than 39.0°C (With cut-off)Air Over Range
Heater power is 100% for more than 15minsChk Sens Position
Heater is unplugged or faultyHeater Open
Heater Power drive fail/uncontrollableChk. Pwr. Drive
Power Fail Alarm: Audio with LED indicator